微囊藻毒素的液-質(zhì)譜聯(lián)用和直接進樣檢測方法比較
發(fā)布時間:2019/11/4 15:42:02 瀏覽次數(shù):1636
微囊藻毒素的液-質(zhì)譜聯(lián)用和直接進樣檢測方法比較
Comparison of Liquid Chromatography Tandem Mass Spectrometry and Direct Injection Methods for the Analysis of Microcystins
DOI:
中文關(guān)鍵詞: 微囊藻毒素 液-質(zhì)譜聯(lián)用 直接進樣 螺螄
英文關(guān)鍵詞: Microcystin liquid chromatography tandem mass spectrometry direct injection analysis Melanioides
采用高效液相色譜-串聯(lián)質(zhì)譜(LC-MS/MS)技術(shù)和直接進樣(DIA)技術(shù)對7種微囊藻毒素進行定性、定量分析,比較兩種方法在精密度、靈敏度、經(jīng)濟性等方面的優(yōu)缺點,并用于螺螄中微囊藻毒素的檢測。LC-MS/MS技術(shù)選用Waters XSelect HSS T3(2.1 mm × 150 mm,3.5 μm),用0.1%甲酸水溶液-0.1%甲酸乙腈溶液作流動相,在電噴霧離子源(ESI)正離子多反應(yīng)離子檢測模式(MRM)下掃描,線性范圍為0.9965~0.9993,相對標準偏差3.6%~6.1%,檢出限和定量限分別小于0.54 ng/mL和1.81 ng/mL,回收率介于73.9%~83.4%之間。DIA技術(shù)選用碎片m/z 135.2為特征母核,優(yōu)化選定DP 60 eV和CE 55 eV,經(jīng)母離子掃描(PreIS)得到各微囊藻毒素的離子峰較為純凈,均以[M+2H]2+為主。兩種方法比較結(jié)果顯示:LC-MS/MS各項參數(shù)均優(yōu)于DIA,而DIA檢測速度較快,僅需1 min即可完成7種微囊藻毒素的快速測定。應(yīng)用兩種方法對60個隨機樣品的檢測表明,螺螄中的微囊藻毒素陽性率達到63.3%,主要污染種類為MC-RR和LR。根據(jù)研究結(jié)果建議當待測樣品數(shù)量較多時,首先采用DIA進行初次篩選,對于陽性樣品利用LC-MS/MS進行二次檢測,可有效減少檢測時間、色譜柱損耗及溶劑消耗,提高結(jié)果的可靠性。本分析策略可用于快速分析水體及動植物中的微囊藻毒素殘留,確定監(jiān)測物種、采樣站位和頻率、警戒值和防控反應(yīng)機制等,為相關(guān)部門完善和強化水產(chǎn)品監(jiān)測和管理工作提供科學(xué)依據(jù),從而保障食用安全和近岸養(yǎng)殖業(yè)的健康發(fā)展。
Seven molecular species of microcystins were qualitatively and quantitatively analyzed by both high performance liquid chromatography-tandem mass spectrometry (LC-MS/MS) and direct injection analysis (DIA). The advantages and disadvantages of these two methods in precision, sensitivity and economy were compared, which were further used for the detection of microcystins in Melanioides. For LC-MS/MS method, the Waters XSelect HSS T3 (2.1 mm × 150 mm, 3.5 μm) was used as separation column with mobile phase of acetonitrile(0.1% formic acid) and water (0.1% formic acid). The MS was operated in electrospray ionization(ESI) and multiple reaction monitoring mode(MRM) with linear range of 0.9965~0.9993 and relative standard deviation of 3.6%~6.1%. The limit of detection and the limit of quantitation were less than 0.54 ng/mL and 1.81 ng/mL, respectively. The recovery ranged from 73.9% to 83.4%. For DIA method, the fragment m/z 135.2 was selected as the characteristic ion with DP 60 eV and CE 55 eV. After precursor ion scanning (PreIS), the microcystin were mainly ionized as [M+2H]2+. The comparison results of these two methods showed that the performance of LC-MS/MS was better than that of DIA, while DIA was faster in detection speed, which only cost one min to complete the determination of seven molecular species of microcystins. Finally, a total of 60 samples was analyzed by the two methods. The results showed that the positive samples were 63.3%, and the main pollution species were Mc-RR and LR. The results suggested that when the number of samples to be tested was large, DIA should be used for initial screening. For the positive samples, LC-MS/MS should be used for the secondary testing, which could effectively reduce the detection time, column loss and solvent consumption, and improve the reliability of the results. This analysis strategy can be applied to fast analysis of microcystins in water, animal, and plant samples, determine the monitoring and sampling position and frequency, the alert value, and the prevention and control, etc. This study could provide scientific basis for related departments to improve and strengthen the monitoring and management of aquatic products. It also could ensure the food safety and the development of offshore aquaculture.